Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (2024)

  • Journal List
  • BMC Musculoskelet Disord
  • v.25; 2024
  • PMC11232251

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (1)

Link to Publisher's site

BMC Musculoskelet Disord. 2024; 25: 526.

Published online 2024 Jul 9. doi:10.1186/s12891-024-07646-0

PMCID: PMC11232251

PMID: 38982393

Yong-De Wu,#1 Xian-Zhong Mei,#2 Wei-Fei Wu,1 Hong-Xi Zhang,3 Jie Liang,1 and Xian-Hua CaiExperimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (2)4,5

Author information Article notes Copyright and License information PMC Disclaimer

Associated Data

Data Availability Statement

Abstract

Background and objective

Complex acetabular fractures involving quadrilateral areas are more challenging to treat during surgery. To date, there has been no ideal internal fixation for these acetabular fractures. The purpose of this study was to evaluate the biomechanical stability of complex acetabular fractures using a dynamic anterior titanium-plate screw system of the quadrilateral area (DAPSQ) by simulating the standing and sitting positions of pelvic specimens.

Materials and methods

Eight formal in-preserved cadaveric pelvises aged 30–50 years were selected as the research objects. First, one hip of the normal pelvises was randomly used as the control model (group B) for measurement, and then one hip of the pelvises was randomly selected to make the fracture model in the 8 intact pelvises as the experimental model (group A) for measurement. In group A, acetabular both-column fractures in the quadrilateral area were established, and the fractures were fixed by DAPSQ. The biomechanical testing machine was used to load (simulated physiological load) from 400N to 700N at a 1mm/min speed for 30s in the vertical direction when the specimens were measured at random in simulated standing or sitting positions in groups. The horizontal displacement and longitudinal displacement of the acetabular fractures in the quadrilateral area were measured in both the standing and sitting simulations.

Results

As the load increased, no dislocation or internal fixation breakage occurred during the measurements. In the standing position, the horizontal displacement of the quadrilateral area fractures in group A and group B appeared to be less than 1mm with loads ranging from 400N to 700N, and there was no significant difference between group A and group B (p > 0.05). The longitudinal displacement appeared to be greater than 1mm with a load of 700mm in group A (700N, 2 cases), and the difference was significant between group A and group B (p < 0.05). In the sitting position, the horizontal and longitudinal displacements of the quadrilateral areas were within 0.5mm in group A and group B, and there was no significant difference between group A and group B (p > 0.05).

Conclusion

For complex acetabular fractures in the quadrilateral area, DAPSQ fixation may provide early sitting stability, but it is inappropriate for patients to stand too early.

Keywords: Acetabular fractures, Quadrilateral area, Articular comminution, DAPSQ biomechanics

Introduction

The term complex acetabular fracture, which involves both columns of the acetabulum, is mainly caused by high-energy injuries [1, 2]. Such fracture patterns, especially in quadrilateral areas, are more challenging to process in trauma departments of orthopedics [3, 4]. Acetabular fractures in the quadrilateral area are usually followed by multiple injuries, open anatomical reduction and internal fixation are traditional therapeutic methods, and the multiapproach operation or modified ilioinguinal approach is adopted during surgery [3, 5, 6]. However, these approaches have many drawbacks, such as nerve and artery injury, easy infection, and slow recovery. Due to complications, a single ilio-inguinal approach has been advocated [7, 8]. Nevertheless, it is difficult to fix the posterior column fracture block (quadrilateral area) through a single anterior approach because the bone in the quadrilateral area is so thin that the screws in this area can easily reach the hip [7, 9]. In recent years, several indirect fixation methods have emerged; for example, lag screws and cerclage wires may produce a certain fixed result [7, 8, 10]. Cerclage wires have been described for managing acetabular fractures in the geriatric population, but they are not commonly used in young patients [10]. However, these fixation methods require more clinical practice and biomechanical analysis, and more work is needed to identify suitable fixators [7, 11]. Wu reported the use of the dynamic anterior titanium-plate screw system of the quadrilateral area (DAPSQ) for the treatment of these fractures, and it produced good clinical effects [8]. The aim of this study was to evaluate the biomechanical stability of the DAPSQ, which would provide early sitting stability after surgery, the standing early maybe inappropriate.

Materials and methods

Specimens

This experimental study was approved by the Ethics Committee of the General Hospital of Central Theater Command of The People’s Liberation Army. In this study, eight 30- to 50-year-old formalin-preserved cadaveric pelvises were selected from Southern Medical University. The specimens with bone abnormalities were excluded by visual observation and X-ray examination. The fourth lumbar vertebra, proximal 1/3 of the femoral shaft and partial soft tissue were preserved in all specimens. To facilitate the measurement, the proximal 1/3 of the femoral shaft was solidified and fixed with denture stone (Fig.1). First, the eight intact pelvises were used as the control models (group B) for measurement, and then, in group A, only one hip of every intact pelvis, where the model of acetabular both-column fracture was randomly established by a wire saw (Fig.2), was used for testing after being fixed by the DAPSQ. The normal hip power in a single-leg stand is approximately 1500N [7], we chose 700N as the ultimate load in the bipedal standing position. After loading, the horizontal displacement of the quadrilateral area in the acetabulum and the longitudinal displacement of the acetabular fractures under 400–700N were recorded. The anatomical reconstruction plate and titanium screws (3.5mm) of the AO were selected for the DAPSQ. A CT machine and X-ray machine (Siemens ARCADIS ORBIC, General Hospital of Central Theater Command of The People’s Liberation Army) were used to assess the safety of the fixation. A ZWICKZ100 material machine (Germany ZWICKZ, a precision of 0.1%, Department of Mechanical Room at Wuhan University of Technology) was used to measure the displacement.

Open in a separate window

Fig. 2

Cadaver fracture model of a high both-column fracture

Fixation system (DAPSQ)

The internal fixation material used was an AO 3.5mm reconstruction plate. The titanium plate of the DAPSQ was slightly upturned and valgus at both ends in the middle section (upper edge of the quadrilateral area) with varus. The degree of bending was approximately 15° (Fig.3a-b). At the same time, the titanium plate was moved to the pelvic cavity at approximately 1/3 to 1/2 of the nail hole on the upper edge of the quadrilateral area, and the plate was screwed although the nail holes were parallel to the surface of the quadrilateral area. More than half of the screws were located in the bone (Fig.3c-f). The auxiliary plate of the iliac wing fracture in the DAPSQ system shares the resistance of the acetabular anterior column plate to displacement of the fracture. Screws in the quadrilateral region of the DAPSQ system may be called semicortical screws. When screws are inserted in the quadrilateral area, the screws are not allowed to break through the inner bone plate (the side bone cortex of the hip joint) under direct observation. We can reshape steel plates, adjust the screw direction or select other screw implants. In this study, for comparative observation, 4 semicortical screws were used in each case of the acetabular area. The length of the semicortical screws should exceed the fracture line by at least 10mm.

Open in a separate window

Fig. 3

Fixation of the dynamic anterior titanium-plate screw system of the quadrilateral area (DAPSQ). (3a-b: Schematic diagram of DAPSQ, 3c: Photo of DAPSQ, 3d-e: CT-ray scan, 3f: X-ray scan of DAPSQ)

Data analysis

Determination of femoral head dislocation: According to the Matta standard [12], anatomical reduction was considered when the fracture displacement did not exceed 1mm after the operation. Therefore, central dislocation of the femoral head was considered if the lateral displacement of the inner wall of the posterior column was greater than 1mm, and upward dislocation of the femoral head was considered if the longitudinal displacement of the anterior inferior spine was greater than 1mm. The specimens were fixed in the position of standing and sitting simulations (Fig.4) in the pelvis as previously described by Pierannunzii et al. and Wu et al., respectively [7, 13]. Before the test, 400N was preloaded to eliminate the effects of bone tissue relaxation, creep, joint space and small differences in fracture reduction, etc., and then loading in the vertical direction was increased from 400N to 700N at a rate of 1mm/min for 30s. According to the simulated physiological weight load, the displacement of the acetabular fractures in the quadrilateral area under the load of 400–700N was measured by the dial gauge of the Zwicks’ 100 electronic universal testing machine in the horizontal and longitudinal directions. The experiment was terminated when hip dislocation or internal fixation rupture occurred.

Open in a separate window

Fig. 4

Testing in the pelvis (4a: standing stimulation, 4b: sitting stimulation)

Statistical analysis

The statistical analysis was conducted using a standard SPSS 26.0 (SPSS Institute, Chicago, IL) software package. The data are expressed as the mean ± standard deviation. The t-tests were performed at each of the 400N, 500N, 600N, and 700N loads to compare the significance of the differences between groups separately for each index of measurement. P < 0.05 was considered to indicate statistical significance, which may indicate that taking the sitting or standing position early after surgery in the DAPSQ system is unsafe.

Results

No dislocation or internal fixation breakage occurred during the measurements. After internal fixation, the acetabular fractures in the quadrilateral area of both columns were immediately fixed, and anatomic reduction was achieved in the A group. There was no screw entering the joint cavity. Soft tissue was taught during the test process. After the removal of the load, the acetabulum and fixation returned to the initial state.

In the standing position (Tables1 and ​and2),2), the loading-horizontal displacements from 400N to 700N and the loading-longitudinal displacements from 400N to 600N appeared to be less than 1mm in group A, the longitudinal displacement appeared to be more than 1mm under 700N loading in group A (700 N2 cases), while the loading-horizontal displacements and the loading-longitudinal displacements appeared to be less than 1mm in group B from 400N to 700N. The statistical analysis revealed that there was no significant difference in the horizontal displacement between group A and group B (p = 0.25), and there was no significant difference in the longitudinal displacement between group A and group B with loads ranging from 400 to 600 (p = 0.14); the difference was significant in the longitudinal displacement between group A and group B with a load of 700 (p = 0.01).

Table 1

Different loading-horizontal displacements in the standing position (Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (7)±s mm)

GroupDifferent loads (N)
400500600700
A0.26 ± 0.080.30 ± 0.140.34 ± 0.180.38 ± 0.23
B0.12 ± 0.020.15 ± 0.030.18 ± 0.070.21 ± 0.11

Open in a separate window

Independent-sample t test: group A - group B (p = 0.25)

Table 2

Different loading-longitudinal displacements in the standing position (Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (8)±s mm)

GroupDifferent loads (N)
400500600700
A0.50 ± 0.160.57 ± 0.240.63 ± 0.250.81 ± 0.39
B0.35 ± 0.140.42 ± 0.200.54 ± 0.220.66 ± 0.21

Open in a separate window

Independent-sample t test: group A - group B (p = 0.14) at each of the 400N, 500N and 600N loads. group A - group B (p = 0.01) at the 700N loads

In the sitting position (Tables3 and ​and4),4), the horizontal and longitudinal displacements of the quadrilateral areas were within 0.5mm between group A and group B. The statistical analysis showed that there was no significant difference between group A and group B (p > 0.05).

Table 3

Different loading-horizontal displacements in the sitting position (Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (9)±s mm)

GroupDifferent loads (N)
400500600700
A0.26 ± 0.080.30 ± 0.090.33 ± 0.100.35 ± 0.11
B0.23 ± 0.070.26 ± 0.080.30 ± 0.130.33 ± 0.17

Open in a separate window

Independent-sample t test: group A - group B (p = 0.65)

Table 4

Different loading-longitudinal displacements in the sitting position (Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (10)±s mm)

GroupDifferent loads (N)
400500600700
A0.20 ± 0.130.24 ± 0.080.27 ± 0.110.31 ± 0.16
B0.11 ± 0.010.15 ± 0.090.18 ± 0.300.21 ± 0.02

Open in a separate window

Independent-sample t test: group A - group B (p = 0.41)

Discussion

The quadrilateral area was located on the inner side of the acetabular posterior column. Quadrilateral area fractures are not clearly defined in the literature, while except for simple fractures according to the classification of Letournel and Judet, the others all involve this area. It is a complex comminuted fracture caused by high-energy injuries under strong trauma from the longitudinal femoral neck. Optimal anatomical reduction and stable fixation of acetabular fractures are important for preventing secondary dislocation and osteoarthritis [14]. We often select a multiapproach operation or a modified ilioinguinal approach for exposure during the operation [12, 15]. Although ORIF has reliable stability, it has limitations in clinical application, such as significant damage due to tissue injury [16]. Single-center series suggest that 12–45% of acetabular fracture patients treated with open reduction internal fixation subsequently undergo at least one secondary surgical procedure that ablates the repaired native acetabulum [17]. Therefore, some authors have claimed that a minimally invasive percutaneous technique should be encouraged, and a single forward approach should be adopted for complex fractures of the acetabulum during surgery [18, 19]. We also found that the ilioinguinal approach provided a good view of fractures of the quadrilateral area and the partial posterior column.

The acetabulum represents a basin-like anatomic structure that bears the body’s weight and accommodates ambulatory activity and hip joint mobility [20]. The pelvic stability provided by the anterior column is 2.75 times greater than that provided by the posterior column [21]. Anatomical reduction in the anterior column combined with rigid internal fixation is the key to preserving hip function via the ilio-inguinal approach.

However, the stability of the hip mainly depends on the blocking effect of the acetabular bone. The bone in the quadrilateral area and the iliopectineal bulge of the anterior column do not enable safe implantation of screws [7, 12]. To avoid this risk, fixations involving no fixation or indirect fixation in quadrilateral areas have emerged. For example, the plate of the anterior column with no fixation in the quadrilateral area only plays a supporting role; for screws away from the acetabular fracture, the degree of fixation stability decreases by approximately 50% [7]. In recent years, drawing support from the development of imaging and navigation technology, a lag screw through the anterior column to the posterior column may produce a certain fixed result [7], but its fixation is eccentric (partial posterior), and it requires that the fracture blocks of the anterior column and posterior column are not crushed [22, 23]. The fracture may be displaced during the tightening of the interfragmentary lag screw. In addition, the cerclage wire, buttress screw or infra-pectineal plates, which are used for elastic fixation, can prevent medial displacement of the medial wall of the acetabulum (quadrilateral area) [2426]. However, the existing hip biomechanics suggest that the bearing force of the acetabulum in the quadrilateral area can be divided into two components in the horizontal and vertical directions in the coronal plane [7]. These kinds of elastic fixation are insufficient for offsetting the vertical separation. Moreover, the stability of buttress plates has always been controversial. A clinical study showed that elastic plate stability was reduced by postoperative turning nursing care, and the plate was easily remodeled to lose its fixation role after fatigue [7, 27].

To our knowledge, there are no direct fixation types in the quadrilateral area in the previous literature, and the effect of bone block in the quadrilateral area is often neglected. Regarding how to improve the stability of the fracture block in the quadrilateral area during the operation, DAPSQ, which uses a single ilioinguinal approach to fix the bone block in the quadrilateral area under direct vision, may be a good choice. Because this is not only the typical case for fixed bone surfaces but also for a small portion of bone, it is dynamic and resistant to separation in the horizontal and vertical directions.

Clinically, how can quadrilateral area screws not enter the joint? The use of repeated intraoperative fluoroscopy or physical examination of hip motion for screening is time-consuming and less reliable. Our operation is carried out visually to avoid screws entering the joint in the DAPSQ system [8], and less intraoperative fluoroscopy is usually used for screening. When screws are inserted in the quadrilateral area, the screws are not allowed to break through the inner bone plate (the side bone cortex of the hip joint) under direct observation. If screws enter the joint, due to the use of reconstruction plates or nonlocking plates, we can reshape the steel plates, adjust the screw direction or select other screw implants. During clinical operation, 3–5 semicortical screws can be selected according to the fracture size of the acetabular area. In our clinical research [8], no screws entered the joint cavity. With the help of the DAPSQ, fractures in the medial acetabular area and the partial posterior column can be fixed perfectly via the classical ilioinguinal approach [8].

To verify the biomechanical stability of the DAPSQ, based on previous studies [7], the selected pelvic specimens were fresh anatomical specimens. All the important ligaments of the pelvic ring were preserved in the specimens. The direction of load conduction was transmitted from the vertebral body to the sacroiliac joint and the acetabulum femoral head. To fully simulate the postoperative physiological status of patients, we selected the sitting and standing positions for the experiment to explore the stability in standing or sitting position.

In group A, because the normal hip power in a single-leg stand is approximately 1500N [7, 28], we chose 700N as the ultimate load in the bipedal standing position. With a maximum load of 700N in the standing position [7], the longitudinal displacement appeared to be more than 1mm, while dislocation of the joint did not occur. This finding is not consistent with the standard proposed by Matta [13]. We suspect that soft tissue may play an important role in restoring joint stability on the DAPSQ. In fracture model designations and measurements, we found that when the fractures were not fixed, the sacrotuberous ligament and the sacrospinous ligament were relaxed, and when the fracture was reduced, especially after loading, the two ligaments recovered tension, which means that the reduction and fixation of the square fracture is closely related to the integrity of the soft tissue. When removing the obturator muscles, especially the sacrospinal ligament and sacral tuberosity ligament, in our experimental fracture model, fixation was prone to instability. We predicted that it would be inappropriate to stand too early with severe soft tissue injury.

In group A, according to measurements of the hip after loading in the simulated sitting position, no dislocation or internal fixation breakage occurred. The maximum displacements of the acetabular area fracture in the horizontal direction and vertical direction were less than 0.5mm. According to the Matta standard [13], anatomical reduction is defined as a fracture displacement that does not exceed 1mm after internal fixation. Perhaps we can infer that it may be safe for patients to sit early after a fracture is fixed according to the DAPSQ.

Although the experiment still offers interesting and valuable data, its limitations must be addressed. Firstly, the sample size is small, we only used eight specimens as representatives because the human pelvis specimen was very precious. Secondly, as eight specimens were tested successively firstly group B and then group A, we chose a relatively small load in this experiment to minimize interference that the changes of the pelvis after the measurement in group B maybe affect the measurement of group A. We ignore the interference to the group B in our study. Thirdly, the observational indicators are insufficient. Due to repeated measurements of specimens, we selected a relatively small limit load in measurements to ensure the integrity of the specimen. Whats more, the acetabular fractures in the quadrilateral area are irregular in shape, the deformation is difficult to measure. For the above reasons, the indexes such as ultimate load and stiffness could not be observed. Fourthly, we did not analyze the role of important soft tissues around the fracture. Moreover, the design control group of this study was the complete pelvis group, and no other internal fixation was used as the internal fixation control group; further comparisons are needed in the future. In addition, the fracture model is relatively simple and does not include serious acetabular fractures for extremely complex quadrilateral fractures in the clinic, especially those with obvious displacement of the posterior column or posterior wall fractures, for which unsatisfactory correction of fracture displacement after anterior approach reduction and fixation is needed to choose an additional posterior approach for fixation.

Conclusion

This preliminary study demonstrated that for complex acetabular fractures in quadrilateral areas, the actual biomechanical effect of DAPSQ was close to that of a normal pelvis on the stability of the hip joint in sitting simulations. DAPSQ fixation may provide early sitting stability, but it is inappropriate for patients to stand too early.

Acknowledgements

The authors would like to thank you. The authors would like to thank the BMC Musculoskeletal Disorders Journal and the editors.

Author contributions

Conception and design: Xian-Hua Cai, Yong-De Wu and Xian-Zhong Mei. Acquisition of data: Yong-De Wu and Xian-Zhong Mei. Analysis and interpretation of imaging data: Yong-De Wu, Xian-Zhong Mei and Wei-Fei Wu. Statistical analysis: Xian-Zhong Mei, Hong-Xi Zhang and Jie Liang. Manuscript Preparation: Yong-De Wu and Xian-Zhong Mei. Manuscript revision and modification: Xian-Hua Cai, Hong-Xi Zhang and Jie Liang. All authors gave final approval of the version to be published, agreed to the submitted journal, and agreed to be accountable for all aspects of the work. These authors contributed equally to this work and should be considered co-first authors: Yong-De Wu and Xian-Zhong Mei.

Funding

The authors do not declare a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

Data availability

The datasets used and analyzed during the current study are available from the corresponding author upon reasonable request. or the Ethics Review Committee of the General Hospital of Central Theater Command of the Chinese People’s Liberation Army.

Declarations

Ethics approval and consent to participate

Informed consent was obtained from the participants’ legal representatives. This report complies with the guidelines for human studies and is in accordance with the Declaration of Helsinki. The ethics review committee of the General Hospital of Central Theater Command of the Chinese People’s Liberation Army General Hospital approved the use of data from these participants in this study.

Consent for publication

Written informed consent was signed by the participants’ legal representative to have the case details and any accompanying images published.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Yong-De Wua and Xian-Zhong Mei contributed equally to this work and should be considered as co-first authors.

References

1. Shaath MK et al. A New Algorithm for Classifying Acetabular Fracture Patterns on Three-dimensional Computed Tomography Reconstructions Markedly Improves Residents’ Ability to Correctly Classify Fractures. J Am Acad Orthop Surg, 2021. [PubMed]

2. Mohan K et al. Acetabular fractures in the elderly: modern challenges and the role of conservative management. Ir J Med Sci, 2021. [PMC free article] [PubMed]

3. Salama W, et al. Modified Stoppa Approach for the treatment of Complex Acetabular fractures. Orthopedics. 2021;44(3):e353–8. doi:10.3928/01477447-20210414-06. [PubMed] [CrossRef] [Google Scholar]

4. Agarwal G et al. Virtual planning on contralateral hemipelvis for posteriorly fixed acetabular fractures. Eur J Trauma Emerg Surg, 2021. [PubMed]

5. Debuka E, et al. Fracture characteristics and outcomes of acetabular fracture management with minimally invasive approach and percutaneous fixation. Eur J Orthop Surg Traumatol. 2021;31(7):1363–8. doi:10.1007/s00590-021-02886-5. [PubMed] [CrossRef] [Google Scholar]

6. Ochs BG, et al. Changes in the treatment of acetabular fractures over 15 years: analysis of 1266 cases treated by the German Pelvic Multicentre Study Group (DAO/DGU) Injury. 2010;41(8):839–51. doi:10.1016/j.injury.2010.04.010. [PubMed] [CrossRef] [Google Scholar]

7. Wu YD, et al. Biomechanical analysis of the acetabular buttress-plate: are complex acetabular fractures in the quadrilateral area stable after treatment with anterior construct plate-1/3 tube buttress plate fixation? Clin (Sao Paulo) 2013;68(7):1028–33. doi:10.6061/clinics/2013(07)22. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

8. Wu H, et al. Single Ilioinguinal Approach to treat Complex Acetabular fractures with quadrilateral plate involvement: outcomes using a Novel Dynamic Anterior plate-screw system. Orthop Surg. 2020;12(2):488–97. doi:10.1111/os.12648. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

9. Giordano V, et al. Associated both-column acetabular fracture: an overview of operative steps and surgical technique. J Clin Orthop Trauma. 2020;11(6):1031–8. doi:10.1016/j.jcot.2020.08.027. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

10. Peter RE. Open reduction and internal fixation of osteoporotic acetabular fractures through the ilio-inguinal approach: use of buttress plates to control medial displacement of the quadrilateral surface. Injury. 2015;46(Suppl 1):S2–7. doi:10.1016/S0020-1383(15)70003-3. [PubMed] [CrossRef] [Google Scholar]

11. Laflamme GY, Hebert-Davies J, Rouleau D. Etal, internal fixation of osteopenic acetabular fractures involving the quadrilateral plate.The injury 2011,42(10):1130–4. [PubMed]

12. Matta JM. Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively with in three weeks after the injury[J] J Bone Joint Surg Am. 1996;78:1632–45. doi:10.2106/00004623-199611000-00002. [PubMed] [CrossRef] [Google Scholar]

13. Pierannunzii L, et al. Acetabular both-column fractures: essentials of operative management. Injury. 2010;41(11):1145–9. doi:10.1016/j.injury.2010.08.011. [PubMed] [CrossRef] [Google Scholar]

14. Hinz N, et al. Biomechanical analysis of fixation methods in acetabular fractures: a systematic review of test setups. Eur J Trauma Emerg Surg. 2022;48:3541–60. doi:10.1007/s00068-022-01936-9. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

15. Mitchell PM, Shaath MK, Routt MJ. Clamp assisted reduction of the transverse Acetabular fracture with anterior column screw fixation and posterior plating: a Technical Trick and Case Series. J Orthop Trauma, 2021. [PubMed]

16. Spalteholz M, Gulow J. Percutaneous triangular stabilization of type 3 and type 4 fragility fractures of the pelvis usually leads to fracture healing despite high revision rates. GMS Interdiscip Plast Reconstr Surg DGPW. 2020;9:Doc05. [PMC free article] [PubMed] [Google Scholar]

17. atterson JT, et al. Early hip survival after open reduction internal fixation of acetabular fracture. Eur J Orthop Surg Traumatol. 2023;33:1209–16. doi:10.1007/s00590-022-03273-4. [PubMed] [CrossRef] [Google Scholar]

18. Wang T, et al. Treatment of acetabular fracture involving anterior and posterior columns using a single pararectus approach: surgical experience and preliminary results. Int Orthop (SICOT) 2023;47:233–40. doi:10.1007/s00264-022-05587-8. [PubMed] [CrossRef] [Google Scholar]

19. Xiao K, et al. Traditional versus mirror three-dimensional printing technology for isolated acetabular fractures: a retrospective study with a median follow-up of 25 months. J Int Med Res. 2021;49(6):3000605211028554. doi:10.1177/03000605211028554. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

20. Brand A et al. Early biomechanical outcome in patients with acetabular fractures treated using the pararectus approach: a gait and stair climb analysis study. Eur J Trauma Emerg Surg, 2021. [PMC free article] [PubMed]

21. Lewis DD, et al. Results of screw/wire/polymethylmethacrylate composite fixation for acetabular fracture repair in 14 dogs. Vet Surg. 1997;26(3):223–34. doi:10.1111/j.1532-950X.1997.tb01489.x. [PubMed] [CrossRef] [Google Scholar]

22. Zhang LH, et al. Experimental study on treatment of acetabular anterior column fractures: applyment of a minimally invasive percutaneous lag screw guide apparatus. BMC Musculoskelet Disord. 2016;17:27. doi:10.1186/s12891-015-0846-1. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

23. Wu X. A biomechanical comparison of different fixation techniques for fractures of the acetabular posterior wall. Int Orthop. 2018;42(3):673–9. doi:10.1007/s00264-017-3728-3. [PubMed] [CrossRef] [Google Scholar]

24. Lin HH, et al. Cerclage wiring in displaced associated anterior column and posterior hemi-transverse acetabular fractures. Injury. 2012;43(6):917–20. doi:10.1016/j.injury.2012.01.016. [PubMed] [CrossRef] [Google Scholar]

25. Meena UK et al. Treatment of acetabular fractures with quadrilateral plate injury-a comparison of two commonly used methods. Orthop Traumatol Surg Res, 2021: p. 102951. [PubMed]

26. Karim et al. Fixation of quadrilateral plate fractures of the acetabulum using the buttress screw: a novel technique[J].INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED,2017,48(8):1813–8. [PubMed]

27. Lei J, et al. Biomechanical analysis of the fixation systems for anterior column and posterior hemi-transverse acetabular fractures. Acta Orthop Traumatol Turc. 2017;51(3):248–53. doi:10.1016/j.aott.2017.02.003. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

28. Graulich T, et al. Posterior acetabular wall morphology is an independent risk factor that affects the occurrence of acetabular wall fracture in patients with traumatic, posterior hip dislocation. Eur J Trauma Emerg Surg. 2023;49:343–9. doi:10.1007/s00068-022-02072-0. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

Articles from BMC Musculoskeletal Disorders are provided here courtesy of BMC

Experimental study on the biomechanical stability of complex acetabular fractures in the quadrilateral area: application of a dynamic anterior titanium-plate screw system (2024)
Top Articles
Appointment Scheduling
Used trucks for sale in the Netherlands - Dutch trucks
Housing near Juneau, WI - craigslist
Mcgeorge Academic Calendar
Metallica - Blackened Lyrics Meaning
Craigslist Cars And Trucks For Sale By Owner Indianapolis
Best Private Elementary Schools In Virginia
Compare the Samsung Galaxy S24 - 256GB - Cobalt Violet vs Apple iPhone 16 Pro - 128GB - Desert Titanium | AT&T
Progressbook Brunswick
Jcpenney At Home Associate Kiosk
Walgreens On Nacogdoches And O'connor
Johnston v. State, 2023 MT 20
Classic Lotto Payout Calculator
Nyuonsite
Colts Snap Counts
Becu Turbotax Discount Code
Leader Times Obituaries Liberal Ks
111 Cubic Inch To Cc
NHS England » Winter and H2 priorities
Odfl4Us Driver Login
Curry Ford Accident Today
What Channel Is Court Tv On Verizon Fios
Walgreens 8 Mile Dequindre
Inbanithi Age
Il Speedtest Rcn Net
Helpers Needed At Once Bug Fables
January 8 Jesus Calling
Random Bibleizer
Jurassic World Exhibition Discount Code
101 Lewman Way Jeffersonville In
Does Royal Honey Work For Erectile Dysfunction - SCOBES-AR
Pickle Juiced 1234
Acadis Portal Missouri
The Boogeyman Showtimes Near Surf Cinemas
Discover Wisconsin Season 16
2023 Fantasy Football Draft Guide: Rankings, cheat sheets and analysis
The best specialist spirits store | Spirituosengalerie Stuttgart
Home Auctions - Real Estate Auctions
Truck Works Dothan Alabama
Timothy Warren Cobb Obituary
The Nikki Catsouras death - HERE the incredible photos | Horror Galore
Chr Pop Pulse
Jimmy John's Near Me Open
Phone Store On 91St Brown Deer
Www Pig11 Net
Haunted Mansion Showtimes Near Millstone 14
Bellelement.com Review: Real Store or A Scam? Read This
Product Test Drive: Garnier BB Cream vs. Garnier BB Cream For Combo/Oily Skin
How to Do a Photoshoot in BitLife - Playbite
Razor Edge Gotti Pitbull Price
Electronics coupons, offers & promotions | The Los Angeles Times
Syrie Funeral Home Obituary
Latest Posts
Article information

Author: Aracelis Kilback

Last Updated:

Views: 5473

Rating: 4.3 / 5 (44 voted)

Reviews: 91% of readers found this page helpful

Author information

Name: Aracelis Kilback

Birthday: 1994-11-22

Address: Apt. 895 30151 Green Plain, Lake Mariela, RI 98141

Phone: +5992291857476

Job: Legal Officer

Hobby: LARPing, role-playing games, Slacklining, Reading, Inline skating, Brazilian jiu-jitsu, Dance

Introduction: My name is Aracelis Kilback, I am a nice, gentle, agreeable, joyous, attractive, combative, gifted person who loves writing and wants to share my knowledge and understanding with you.